Date: Thu, 10 Nov 94 16:27:13 CST
Subject: SX50 green monitors
From:
Status: OR
Date: Thu, 10 Nov 1994 15:38:01 PST
From: "Dr. Claudio Cermignani"
Subject: SX50 green monitors
The monitors' scanning is controlled by the SX50. As for the folded over
images, the problem appears to be with the monitor's power supply
(located in the backside of the monitor) to the X scanning. Check the two
regulators on the X-scanning and replace both, even if only one is shot.
Check 15V and 20V lines. Good luck.
Claudio Cermignani
Dept of Geology
Microprobe Lab
University of Toronto
-----------
Date: Thu, 10 Nov 94 13:27:24 PST
From: "michael shaffer"
Subject: Re: SX50 crt info requested
I believe it is independent, at least the power drivers are local to
the monitor. They may be under SX control, but the problem is local. I
have the same problem, and we have been unable to fix it. We
troubleshot it at the time by "freezing" one of the high power
consuming xsistors, but replacing it helped only short term. If you
find a fix let me know.
--Michael Shaffer